Easyfield
Service manual

Questions and Answers

  • W
    wjordanJul 29, 2025
    Why is the CCD picture dark on my OCULUS Medical Equipment?
    • I
      Isaiah AndersenJul 29, 2025
      If the CCD picture on your OCULUS Medical Equipment appears dark, it could be due to several reasons. First, the brightness might not be set correctly, so you should adjust the CCD brightness. Another possible cause is a defect in the infrared LED used to illuminate the patient's eye, in which case you should replace the infrared LED. Finally, the CCD chip itself might be defective, requiring you to replace the CCD chip.
  • L
    Logan HawkinsAug 3, 2025
    What to do if one of the hand held units control buttons does not work on OCULUS Medical Equipment?
    • S
      shawn28Aug 3, 2025
      If one of the control buttons on your OCULUS Medical Equipment hand-held unit isn't working, the button itself might be defective. In this case, replace the faulty button with a new one.
  • J
    james24Aug 7, 2025
    How to fix a blur CCD picture on OCULUS Medical Equipment?
    • R
      Randy JordanAug 8, 2025
      If the CCD picture on your OCULUS Medical Equipment is blurry, it might be because the CCD board isn't set up correctly. To resolve this, install the CCD board correctly.
  • N
    Nicole McfarlandAug 13, 2025
    What to do if CCD picture is disoriented on OCULUS Easyfield?
    • D
      davidmeyerAug 13, 2025
      If the CCD picture on your OCULUS Medical Equipment looks disoriented, the CCD board might not be set up correctly. To fix this, install the CCD board correctly.
  • J
    Jordan OwensAug 17, 2025
    Why does the testpoint LED not illuminate on OCULUS Medical Equipment?
    • M
      Molly AndersonAug 17, 2025
      If the test point LED on your OCULUS Medical Equipment doesn't illuminate, the LED itself is likely defective and needs to be replaced. Be careful! The LED has to be recalibrated after replacement!
  • U
    ujonesAug 22, 2025
    How to fix a chessboard-like CCD picture with stripes on OCULUS Easyfield?
    • M
      Matthew SmallAug 22, 2025
      If the CCD picture on your OCULUS Medical Equipment looks like a chessboard or has vertical or horizontal stripes, the CCD board is likely defective and needs to be exchanged.

Summary

1 Safety Precautions

General Safety Guidelines

Covers protection from environmental factors, proper usage, and user training requirements for the measuring instrument.

3 System Requirements

PC Specifications

Details minimum PC hardware and operating system requirements for controlling the Easyfield device.

4 System Setup

PC Connection and Wiring

Instructions on connecting the Easyfield to a PC via RS232 and connecting the power supply unit.

5 Basic Operation Principle

System Workflow and Components

Describes how the system illuminates the eye, captures images via CCD camera, and processes data using a microcontroller.

6 Easyfield Device Operations

7 Troubleshooting Guide

Common Malfunctions and Remedies

Addresses issues like incorrect image display, dark pictures, or non-functional buttons with their possible causes and solutions.