What to do if OHAUS Scale is busy?
- GGregory RodriguezSep 23, 2025
Wait until completion.
What to do if OHAUS Scale is busy?
Wait until completion.
How to fix OHAUS AV412 that does not display accurately?
If your OHAUS Scales are not displaying accurately, it could be due to improper calibration. Try performing a calibration. Another possible cause is an unstable environment. Move the balance to a more suitable, stable location.
Why I cannot calibrate the OHAUS AV412?
If you cannot calibrate your OHAUS Scales, it could be due to several reasons: 1. The calibration menu might be locked; turn the calibration menu lock off. 2. LFT might be set to on; turn LFT off. 3. There might be an unstable weight reading; eliminate vibration and drafts.
How to fix OHAUS Scale that will not turn on?
If your OHAUS Scale won't turn on, first ensure the AC adapter is properly connected. If it still doesn't power up, the batteries may be discharged and need replacing.
What to do if battery indicator is flashing on OHAUS AV412?
If the battery indicator is flashing on your OHAUS Scales, it means the batteries are weak and need to be replaced.
What does error 8.4 mean on OHAUS Scale?
Error 8.4 on your OHAUS Scale means the weight reading is below the underload limit. Place the platform on the balance.
What does error 8.1 mean on OHAUS Scale?
Error 8.1 on your OHAUS Scale means the weight reading exceeds the Power On Zero limit. Make sure you clear the platform.
What does error 8.2 mean on OHAUS Scale?
Error 8.2 on your OHAUS Scale indicates the weight reading is below the Power On Zero limit. Ensure the platform is properly placed on the balance.
What does error 9.8 mean on OHAUS AV412 Scales?
Error 9.8 on OHAUS Scales indicates that user calibration data is not present, which is required when LFT is turned on. Calibrate the balance to resolve this.
What does Error 8.4 mean on OHAUS AV412?
Error 8.4 on your OHAUS Scales indicates that the weight reading is below the underload limit. To fix this, put the platform on the balance.
| Capacity | 410 g |
|---|---|
| Readability | 0.01 g |
| Calibration | External |
| Display | LCD |
| Linearity | ±0.02 g |
| Stabilization Time | 2 seconds |
| Interface | RS232 |
| Power Supply | AC Adapter |
| Units of Measurement | g, oz, ozt, dwt, ct |
| Operating Temperature Range | 10°C – 40°C |