racheljenningsAug 20, 2025
What to do if OHAUS Scales show system errors?
- CCindy PrestonAug 20, 2025
Turn the power off, then turn it back on. If the error persists, have the balance serviced.
What to do if OHAUS Scales show system errors?
Turn the power off, then turn it back on. If the error persists, have the balance serviced.
What to do if OHAUS AP210 Scales show error 8.3, hardware error caused by an internal weight signal which is too high?
Check the load on the platform which may be excessive. If the error persists, the balance must be serviced.
What does error 4.4 mean on OHAUS AP210?
Error 4.4, RS232 buffer is full, may occur if no printer or computer is connected to the interface. To clear the buffer, turn the balance off or enter the Print menu and select END.
What to do if OHAUS Scales show error 8.2, internal power on error?
Turn the power off, then turn it back on.
How to fix OHAUS AP210 that will not turn on?
If your OHAUS Scales are not turning on, the power cord may not be plugged in or properly connected to the balance. Check the power cord connections to ensure they are secure.
How to store menu settings on OHAUS AP210?
To save menu settings, you MUST use End to exit menus.
What does error 9.7 mean on OHAUS Scales?
Error 9.7, Invalid setup data checksum. Check Setup, User, and Print menus settings. If possible, try to enter menus and exit using END to restore menu settings. It may be caused by a faulty component, or in rare cases, a severe static charge. If the error persists, the balance must be serviced.
What does error 2.0 mean on OHAUS Scales?
Error 2.0, Illegal tare operation or balance is unable to stabilize within time limit after taring, may happen because the environment is too hostile or the balance needs recalibration. Recalibrate the balance.
What does error 3.0 mean on OHAUS Scales?
Error 3.0, Incorrect or no calibration weight used for User calibration. Recalibrate with correct weights.
What does error 5.0 mean on OHAUS AP210?
Error 5.0, Auto calibration failed, may happen because the environment is too hostile. Recalibrate the balance.
Capacity | 210 g |
---|---|
Pan Size | Ø 90 mm |
Display | LCD |
Calibration | External calibration |
Housing Material | ABS plastic |
Interface | RS232 |
Readability | 0.001g |
Power Supply | AC adapter |
Stabilization Time | 3 seconds |
Units of Measurement | g, oz, ct, dwt, tl |
Procedure for measuring the weight of an object on the pan.
Steps for zeroing out the weight of a container before weighing.
Explains Auto, User, and Test calibration methods and protection.
How to adapt the balance to environmental conditions and set parameters.
Guides on customizing balance operations, units, and functions.
Covers reset, communication, GLP, and print options settings.