Questions and Answers

  • A
    Amanda DoyleAug 20, 2025
    What to do if OHAUS Scales show system errors?
    • J
      Jacqueline RossAug 20, 2025
      Turn the power off, then turn it back on. If the error persists, have the balance serviced.
  • C
    carolhernandezAug 22, 2025
    What to do if OHAUS AP250D show error 8.3, hardware error caused by an internal weight signal which is too high?
    • M
      Michael BakerAug 22, 2025
      Check the load on the platform which may be excessive. If the error persists, the balance must be serviced.
  • J
    johncabreraAug 25, 2025
    What does error 4.4 mean on OHAUS AP250D Scales?
    • L
      Laura BowmanAug 25, 2025
      Error 4.4, RS232 buffer is full, may occur if no printer or computer is connected to the interface. To clear the buffer, turn the balance off or enter the Print menu and select END.
  • S
    Sherri LopezAug 26, 2025
    What to do if OHAUS AP250D Scales show error 8.2, internal power on error?
    • Z
      znguyenAug 26, 2025
      Turn the power off, then turn it back on.
  • E
    Eduardo AndersonAug 27, 2025
    How to fix OHAUS Scales that will not turn on?
    • S
      Sharon FrazierAug 27, 2025
      If your OHAUS Scales are not turning on, the power cord may not be plugged in or properly connected to the balance. Check the power cord connections to ensure they are secure.
  • A
    alyssaloweAug 29, 2025
    How to store menu settings on OHAUS AP250D?
    • P
      Patricia MelendezAug 30, 2025
      To save menu settings, you MUST use End to exit menus.
  • A
    awellsAug 31, 2025
    What does error 9.7 mean on OHAUS AP250D Scales?
    • L
      Luke BlairAug 31, 2025
      Error 9.7, Invalid setup data checksum. Check Setup, User, and Print menus settings. If possible, try to enter menus and exit using END to restore menu settings. It may be caused by a faulty component, or in rare cases, a severe static charge. If the error persists, the balance must be serviced.
  • M
    Melissa PierceSep 2, 2025
    What does error 2.0 mean on OHAUS Scales?
    • K
      Kimberly HallSep 2, 2025
      Error 2.0, Illegal tare operation or balance is unable to stabilize within time limit after taring, may happen because the environment is too hostile or the balance needs recalibration. Recalibrate the balance.
  • C
    Clayton BensonSep 4, 2025
    What does error 3.0 mean on OHAUS Scales?
    • A
      Andrew TorresSep 4, 2025
      Error 3.0, Incorrect or no calibration weight used for User calibration. Recalibrate with correct weights.
  • J
    john17Sep 6, 2025
    What does error 5.0 mean on OHAUS AP250D?
    • P
      Pamela NelsonSep 6, 2025
      Error 5.0, Auto calibration failed, may happen because the environment is too hostile. Recalibrate the balance.

Summary

Operating the Balance

Navigating Balance Menus

Guide on how to access and move through the balance's menu system.

Core Weighing Modes

Covers weighing, taring, percent, parts counting, check weighing, animal weighing, and fill guide.

Menus and Configuration

Calibration Menu Options

Details on auto, user, and test calibration methods.

Maintenance and Support

Troubleshooting Common Issues

Common problems and their solutions for balance operation.

Error Codes and Remedies

Lists various error codes and their corresponding remedies.

Limited Warranty Information

OHAUS AP250D Specifications

General IconGeneral
Capacity250 g
Pan Size90 mm
DisplayLCD
Stabilization Time2 seconds
Power SupplyAC Adapter
Units of Measurementg, oz
InterfaceRS232
Operating Temperature15°C to 30°C
CalibrationExternal Calibration

Related product manuals