STARTER 3100M Bench Meter EN-3
1.3 Display and controls
Displays
Slope: more than 95%
and offset: ± (0-15) mV
Electrode condition is good
Slope: 90-95%
or offset: ± (15-35) mV
Electrode condition is
acceptable
Slope: less than 90%
or offset: ± (35-60) mV
Electrode condition is not
good or needs cleaning
Endpoint stability icon ; Auto endpoint icon
.
Measurement icon - ; means measurement or calibration is running when blinks.
Calibration icon - ; means calibration in progress when display.
Setup icon - ; instrument is in the setup mode, can set temperature(MTC), buffer group
etc.
pH/mV reading, or slope of pH electrode calibration; or Conductivity/TDS/Salinity/Resistivity
value, or cell constant of conductivity electrode.
Calibration point / Buffer group /Memory number / Error index
.
Auto temperature compensation - ATC ; Manual temperature compensation – MTC;
Conductivity reference temperature - R.Temp
Temperature value during measurement or offset (mV) value in pH calibration process.