STARTER 3100M Bench Meter EN-3
1.3 Display and controls
Displays
1
pH Electrode condition
Slope: more than 95%
and offset: ± (0-15) mV
Electrode condition is good
Slope: 90-95%
or offset: ± (15-35) mV
Electrode condition is
acceptable
2
Endpoint stability icon ; Auto endpoint icon
.
3
Measurement icon - ; means measurement or calibration is running when blinks.
4
Calibration icon - ; means calibration in progress when display.
5
Setup icon - ; instrument is in the setup mode, can set temperature(MTC), buffer group
etc.
6
pH/mV reading, or slope of pH electrode calibration; or Conductivity/TDS/Salinity/Resistivity
value, or cell constant of conductivity electrode.
7
Calibration point / Buffer group /Memory number / Error index
.
8
Auto temperature compensation - ATC ; Manual temperature compensation – MTC;
Conductivity reference temperature - R.Temp
9
Temperature value during measurement or offset (mV) value in pH calibration process.
2
3
4
1
5
7
8
9
1
2
3
4
5
6