Instructions for Use EN Olympus AFU-100
24
WB135745-W10
DANGER
If inadequate flow is observed or the flow does not stop, end the
procedure immediately. Use the power switch (0) as emergency stop for
malfunctions (e.g. in case the flow activation / deactivation button (4) or
the footswitch does not react).
The peristaltic pump unit has no suction function! Keep a suction pump
available in case excessive liquid must be removed.
If a flow level is set below 20, the peristaltic pump unit does not stop
automatically. Unintended amounts of liquid can accumulate within the
patient’s body.
WARNING
Wear personal protective equipment to guard against dangerous
chemicals and potentially infectious material. During operation, wear
appropriate personal protective equipment, such as eye wear, face
mask, moisture-resistant clothing and chemical-resistant waterproof
gloves that fit properly and are long enough so that your skin is not
exposed.
Check the flow level before each application. The physician has the
responsibility to verify the effect of the flow level setting.
Never use toxic, acid or non-sterile liquids. Otherwise, it may cause
user and / or patient toxications, chemical burns or infections.
If there is a malfunction at the flow activation / deactivation button (4) or
the footswitch pedal (21) or the electrosurgical unit’s footswitch,
continuous flow may cause patient injury.
NOTE
If the peristaltic pump unit is connected with the footswitch via the
footswitch plug (22), the footswitch assignment indicator (13) illuminates
and activation / deactivation is possible with the footswitch pedal (21). If
the peristaltic pump unit is connected with an electrosurgical unit, the
electrosurgical unit indicator (12) illuminates and
activation / deactivation is possible with a special button on the
electrosurgical unit’s footswitch. The flow activation / deactivation button
(4) can always be used to activate or deactivate the flow.
To avoid patient injury, the maximum activation time is limited to 20
seconds, if a flow level higher than 20 has been set. After exceeding the
time limit, the unit will be deactivated automatically. To continue the
procedure, the flow has to be activated again.
7.7 Procedure after use
1) Switch off the peristaltic pump unit.
2) Remove the pump tube.
3) If the peristaltic pump unit is not used for a long time period, disconnect the power
cord plug from the grounded wall outlet.
4) Clean the components following the instructions in chapter 8 (Cleaning, storage
and disposal).
CAUTION
When disconnecting plugs of accessories or power cords, always hold
the plug. Pulling the cable may result in damaging of the wires.