Chapter 1 General Policy
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
When reprocessing an endoscope that has the flexibility
adjustment mechanism, make sure that the insertion tube is
set to the softest condition (indicated by the “” mark on the
flexibility adjustment ring) before immersing the endoscope in
cleaning/disinfectant solution.
If the endoscope is cleaned and disinfected or sterilized while
the insertion tube is stiff, the endoscope may be damaged.
Endoscope models Water-resistant cap
CF-Q160ZL/I MH-553, MAJ-583
CF-Q160DL/I MH-553, MAJ-942
Other MH-553
Table 1.1