Chapter 3 Cleaning, Disinfection and Sterilization Procedures
37
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
9. Wait 30 seconds, or until the covering of the bending section contracts to its
pre-expansion size. Disconnect the leakage tester’s connector cap from the
venting connector.
10. Thoroughly dry the leakage tester.
3.5 Manual cleaning
After completing the leakage test, perform manual cleaning according to the
procedures described below.
Equipment needed
Prepare the following equipment:
To prevent damage to the endoscope, never immerse it
together with objects other than the reprocessing
accessories.
Personal protective equipment
Soft brush
Clean, lint-free cloths
Large basins
Detergent solution
Clean water
Suction pump
•30cm
3
(30 ml) syringe
Suction cleaning adapter (MH-856)
Channel plug (MH-944)
Injection tube (MH-946)
Channel cleaning brush (BW-20T)
Channel cleaning brush (BW-17K for GIF-XTQ160 only)
Channel-opening cleaning brush (MH-507)
Auxiliary water tube (MAJ-855 for endoscopes with
auxiliary water feeding only)