Chapter 3 Cleaning, Disinfection and Sterilization Procedures
45
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
7. Attach the 30 cm
3
(30 ml) syringe to the injection tube’s air/water channel
port (see Figure 3.27).
8. Inject 90 cm
3
(90 ml) of detergent solution into the air/water channel.
9. Disconnect the channel plug and injection tube from the endoscope, and
leave all items immersed.
Figure 3.27
Biopsy valve cap
Channel plug
Air channel
Air/water cylinder
Suction cylinder
Suction connector
Suction port
Suction channel port
Syringe
Water channel
Air/water nozzle
Connector plug
Water supply connector
Air supply connector
Air pipe
Injection tube
Air/water channel port
Air pipe port