TROUBLESHOOTING CLV-160
Lamp check
The light source xenon lamp is not lit.
Power supply
check
Panel operation
check
Does the audible
warning ring?
Yes
Check the temperature
switch.
• ON if normal.
No
No
Does the audible
warning ring?
Yes
Replace the UPCL16CR00.
(Replace the main board.)
Does the xenon
lamp light up?
No
Yes
Check the lamp mounting.
The lamp is correctly mounted.
The heat compound is correctly
painted.
Yes
Does the xenon
lamp light up?
No
Replace the lamp by a
brand-new one (Y1064S).
Yes
Yes
Does the xenon
lamp light up?
No
Check around the lamp
housing.
Does the xenon
lamp light up?
Unnecessary things (metallic pieces, etc.)
High voltage pulse leaks.
Connection with the electrode F and
electrode R.
The round terminal is close to the lamp
housing.
No
A
B
4-4