TROUBLESHOOTING CLV-160
Emergency lamp
operation check
Panel operation
check
Is the emergency
lamp LED lit on the
front panel after the
power is turned on?
Yes
No
The emergency lamp is not automatically lit
when the xenon lamp goes out.
No
No
Replace the emergency
lamp.
Is the emergency
lamp LED lit on the
front panel?
Yes
Replace the W12M018A.
(Replace the emergency
lamp harness.)
Is the emergency
lamp LED lit on the
front panel?
Yes
Does the emergency
lamp state signal
detect the defect?
Yes
No
J4/7 pin of the main
board is H.
Replace the UPCL16PS10/20.
Replace the UPCL16CR00.
(Replace the switching power
(Replace the main board.)
Is the emergency lamp
lit when the main lamp
is not lit?
Yes
No
Remove the xenon
lamp, and make the
emergency lamp
light up.
Is the turret plate
rotated to the
emergency lamp
position?
Yes
Replace the
emergency lamp.
No
A
B
C