TROUBLESHOOTING CLV-160
Transillumination
Brightness does not change even if the
function check
transillumination switch is depressed.
Panel operation
check
Filter switching
operation check
Manual brightness
adjustment check
Is the brightness
increased to the maximum
when the
transillumination switch
is depressed?
No
Yes
Replace the UPCL16CR00.
(Replace the main board.)
End