DISASSEMBLING PROCEDURE CLV-160
1.
5. DISASSEMBLING PROCEDURE
General Precautions on Disassembling
Replace the parts and wires to the original positions.
For electrical safety and standard, be sure to reassemble the following parts to the original
states.
1. Insulation tube and mylar sheet
2. Cables clamped and separated from the heating parts or high-voltage parts
3. Cover screws with a toothed lock washer to suppress a radiation noise
Use the specified parts.
The parts used in this unit are designed protective against vibration, heat and high voltage. Be
sure to select the parts with the same characteristics from the parts list when replacing the
parts.
Be careful when disconnecting the cable housing.
Don’t pull the cable. Be sure to use the special tool.
2. Jigs and Tools
No
Name
1
Phillips
screwdriver No.
2
2
Box screwdriver(5.5mm)
3
Nippers
No.
OT0287
OT0078
OT0051
Specification
150mm
Remarks