EasyManua.ls Logo

Olympus Delta - Page 9

Olympus Delta
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
103158-01EN [U8998465], Rev. A, April 2012
Configuring the Workstation 9
Figure 5-3 Test-stand locking fixture
3. Align the large tab of the probe adaptor with the large notch of the locking fixture (see Figure 5-4 on page 9).
Figure 5-4 Alignment with locking fixture
4. Make sure that the probe-adaptor faceplate is flush with the test chamber floor.
5. Carefully turn the analyzer-adaptor assembly counterclockwise until it locks.
The probe-adaptor plastic housing must be centered with the locking-fixture alignment screw (see Figure 5-5 on
page 9).
Figure 5-5 Analyzer-adaptor assembly installed on the test stand
6. Place the workstation in the upright position
Locking fixture
Locking-fixture large notch
Probe-adaptor large tab
Alignment screw
Plastic housing

Related product manuals