48
DP74
Problem Cause Remedy Page
The 5760 x 3600, 2880 x 1800
and 1920 x 1200 images are not
clear.
The camera is subjected to vibration
during image acquisition.
Acquire images in an environment
where the microscope and camera
are not vibrated. It is effective to use an
anti-vibration table.
13
The temperature inside the camera
head is not stabilized.
Wait 2 minutes or more after starting
the software until the temperature
inside the camera head is stabilized,
and acquire the image.
The files acquired and saved
with 10-bit color/10-bit grayscale
cannot be opened properly by
the software other than cellSens
/ OLYMPUS Stream. Or, the
image in black color is displayed.
(including the icon on the window)
The effective data is lower 10-bit out of
16-bit file format. However, your software
does not support 16-bit format or
displays higher 8-bit out of 16-bit.
Use cellSens / OLYMPUS Stream to
open the image file.
The window of DP2-TWAIN,
cellSens or OLYMPUS Stream
is not displayed properly. Or
characters of the menu are not
displayed properly.
The resolution of the display is not set
appropriately.
Set the resolution of the display to 1280
x 1024 or higher.
The "large font" is selected as the font
size of the display.
Select the "small font" as the font size
of the display.