Chapter 2 Instrument Nomenclature and Specifications
Rotatable Clip Fixing Device
12
Operating Environment
Specifications
Ambient Temperature 10 to 40°C (50 to 104°F)
Relative Humidity 30 to 85%
Air Pressure 700 to 1060 hPa
(0.71 to 1.08 kgf/cm
2
)
(10.1 to 15.4 psia)
Model HX-110LR HX-110QR
Shape of the
distal end of
the insertion
portion
Maximum
insertion
portion
diameter
(mm)
ø 2.75
Working
length (mm)
1650 1950
Compatible
Olympus
endoscopes
(All of these
parameters
should be met.)
Length and
model
Working
length
less than 1200
mm;
EF, GIF, OGF,
CF (Exclude I-,
L-length), OSF
Working
length
less than 1500
mm;
EF, GIF, OGF,
CF (Exclude
L-length),
PCF (I-length
only), OSF
Channel inner
diameter
(mm)
(Color code)
ø 2.8, ø 3.2
(Yellow);
ø 3.7, ø 6
(Orange)
ø 2.8, ø 3.2
(Yellow);
ø 3.7, ø 4.2,
ø 6 (Orange)