Rotatable Clip Fixing Device
25
Chapter 3 Preparation, Inspection and Operation
Do not insert the instrument into the
endoscope unless you have a clear
endoscopic field of view. If you cannot
see the distal end of the insertion portion
in the endoscopic field of view, do not
use it. This could cause patient injury,
such as perforation, bleeding or mucous
membrane damage. It may also damage
the endoscope, instrument and/or clip.
Do not angulate the bending section of
the endoscope abruptly while the distal
end of the insertion portion is extended
from the distal end of the endoscope.
This could cause patient injury, such as
perforation, bleeding or mucous
membrane damage.
Do not force the distal end of the
insertion portion against body cavity
tissue. This could cause patient injury,
such as perforation, bleeding or mucous
membrane damage.
When electrosurgical accessories are
used after clipping, they could cause
patient injury, such as thermal injury of
body cavity tissue contacting the clip.
Activate output only after checking the
tissue around the clip.
When using the instrument with a two
channel endoscope, never use
electrosurgical accessories at the same
time. This could cause patient, operator
or assistant injury, such as thermal
injury.