Rotatable Clip Fixing Device
29
Chapter 3 Preparation, Inspection and Operation
Insertion Into the Endoscope
Do not force the instrument if resistance
to insertion is encountered. Reduce the
angulation of the endoscope until the
instrument passes smoothly. Attempting
to force the instrument could cause
patient injury, such as perforation,
bleeding or mucous membrane damage.
It may also damage the endoscope and/
or instrument.
Hold the slider still when inserting the
instrument into the endoscope.
Otherwise, the clip will open and may
extend from the distal end of the
endoscope abruptly. This could cause
patient injury, such as perforation,
bleeding or mucous membrane damage.
It could also damage the endoscope,
instrument and/or clip.
Do not advance or extend the instrument
abruptly. This could cause patient injury,
such as perforation, bleeding or mucous
membrane damage. It could also
damage the endoscope, instrument and/
or clip.
When inserting the instrument into the
endoscope, make sure that the clip is
completely retracted into the coil sheath.
Otherwise, patient injury, such as
perforation, bleeding or mucous
membrane damage may result. It could
also damage the endoscope, instrument
and/or clip.