10
Important Information — Please Read Before Use
EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL
Do not touch the electrical contacts inside the electrical
connector. CCD damage may result.
Do not apply shock to the distal end of the insertion section,
particularly the objective lens surface at the distal end. Visual
irregularities may result.
Do not twist or bend the bending section with your hands.
Equipment damage may result.
Do not squeeze the bending section forcefully. The covering
of the bending section may stretch or break and cause water
leaks.
Turn the video system center ON only when the videoscope
cable is connected to both the video system center and the
electrical connector on the endoscope. In particular, confirm
that the video system center is OFF before connecting or
disconnecting the videoscope cable from the electrical
connector on the endoscope. Failure to do so can result in
equipment damage, including destruction of the CCD.
The endoscope’s remote switches cannot be removed from
the control section. Pressing, pulling, or twisting them with
excessive force can break the switches and/or cause water
leaks.
If remote switch 1 does not return to the OFF position after
being pressed strongly from the side, gently pull the switch
upwards to return it to the OFF position.
Do not hit or bend the electrical contacts on the endoscope
connector. The connection to the light source may be
impaired and faulty contact can result.
Do not pull the magnification control cable during an
examination. The endoscope connector will be pulled out
from the output socket of the light source and the endoscopic
image will not be visible. The magnification control cable will
be pulled out from the endoscope’s zoom connector and the
endoscopic image will not zoom in or out (for CF-H260AZL/I
only).
Turn the magnification controller OFF before connecting or
disconnecting the magnification control cable from the
endoscope’s zoom connector. Failure to do so can result in
equipment damage (for CF-H260AZL/I only).