Important Information — Please Read Before Use
11
EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL
Do not pull the UPD cable during an examination. The
endoscope connector will be pulled out from the output
socket of the light source and the endoscopic image will not
be visible. The UPD cable will be pulled out from the UPD
scope connector and the scope model will not be visible (for
CF-Q260DL/I, CF-H260DL/I only).
Turn the endoscope position detecting unit (UPD) OFF
before connecting or disconnecting the UPD cable from the
UPD scope connector. Turn the UPD ON or OFF only when
the UPD cable is connected to the UPD scope connector.
Failure to do so can damage the UPD (for CF-Q260DL/I,
CF-H260DL/I only).
Electromagnetic interference may occur on this instrument
near equipment marked with the following symbol or other
portable and mobile RF (radio frequency) communications
equipment, such as cellular phones. If electromagnetic
interference occurs, mitigation measures may be necessary,
such as reorienting or relocating this instrument, or shielding
the location.
Be sure that this instrument is not used adjacent to or
stacked with other equipment (other than the components of
this instrument or system) to avoid electromagnetic
interference.
To check the electromagnetic interference from other
equipment (any equipment other than this instrument or the
components that constitute this system), the system should
be observed to verify its normal operation in the configuration
in that it will be used.