76
Chapter 3 Preparation and Inspection
EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL
The NBI observation mode is available when the video
system center CV-260SL and the light source CLV-260SL,
CLV-260NBI are used with the endoscope. The CLV-260NBI
may not be available in some areas.
Connection of the endoscope and ancillary equipment
Firmly connect the suction tube from the suction pump to the
suction connector on the endoscope connector. If the suction
tube is not attached properly, debris may drip from the tube
and can pose an infection control risk, cause equipment
damage, and/or reduce suction capability.
The GIF-H260, GIF-H260Z, CF-H260AL/I, CF-H260AZL/I,
and CF-H260DL/I can only be connected to the video system
center CV-260SL, CV-260.
1. If any ancillary equipment is ON, turn it OFF.
2. Insert the endoscope connector completely into the output socket (scope
socket when using the CLV-260) of the light source.
3. Place the water container’s water supply channel onto the water supply
connector on the endoscope connector at an angle of 90 and push it in until
it stops (see Figure 3.20 (1)).
4. Turn the water container’s connection adapter 90 clockwise to align the air
supply channel with the air supply connector on the endoscope connector
(see Figure 3.20 (2)).
5. Push the water container’s connection adapter again until it stops (see
Figure 3.20 (3)).
6. Confirm that the water container’s connection adapter fits properly and that
it cannot be rotated (see Figure 3.20 (4)).