Chapter 3 Cleaning, Disinfection and Sterilization Procedures
63
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
1. Seal the individual parts or equipment separately in packages appropriate
for steam sterilization (autoclaving) according to your hospital’s protocol.
2. Steam sterilize the packages according to the recommended steam
sterilization (autoclaving) exposure parameters given in Section 2.6, “Steam
sterilization (autoclaving) of accessories” and the sterilizer manufacturer’s
instructions.
3. Following steam sterilization (autoclaving), let all components gradually cool
down to room temperature.
4. Store the components following the instructions given in chapter 5, “Storage
and Disposal”.
3.10 Cleaning, disinfection and sterilization
procedures for distal cover
Equipment needed
Prepare the following equipment, and wear appropriate personal protective
equipment.
Personal protective equipment
Small basin
Clean water (non-sterile water)
Detergent solution
Soft brush
Clean lint-free cloths
•30cm
3
(30 ml) syringe
Disinfectant solution
Sterile water for sterile water rinse
Sterile lint-free cloths
70% ethyl or isopropyl alcohol