6 Hysteroflow Quick Reference Guide
Warnings
Begin the procedure
Begin the Procedure
1. Press
Details
to display inflow
and outflow volumes.
2. Press (–) or (+) to set Nominal Pressure.
3. Press
Reset Actual Deficit
, then
press
Yes
.
4. Open inflow then open outflow
stopcocks on the instrument to
begin the procedure.
Warning Message Alert Cause
Change Fluid Bag
3 Beeps > 500 ml remaining
Deficit Reached
3 Beeps Deficit exceeds the preset value
Overpressure
3 Beeps Actual pressure 10 mmHg above nominal pressure
Deficit Exceeded
10 Beeps In increments of 100 ml above the preset value
Critical Overpressure
10 Beeps
Actual pressure > 200 mmHg above nominal
pressure
Rapid Deficit
10 Beeps Deficit change > 300 ml in 1 minute