PolyLoop Ligating Device
(
HX-400U-30
)
Quick Reference Guide
Once in place, pull the yellow cylinder
proximally until it stops to extend the
PolyLoop from the plastic sheath.
Remove fingers from the yellow cylinder.
Position the PolyLoop over the polyp at the base
of the stalk, approximately 1-2 mm above the
mucosa. Pull the slider towards the thumb ring
to ligate the tissue. Look at the screen and feel
the resistance in your hand to ensure you do not
transect the polyp stalk. Caution: Do not apply
unnecessary force to the loop when ligating
tissue during the procedure.
With the yellow cylinder still touching the
handle, push the slider to release the
PolyLoop.
If the PolyLoop gets stuck in the tube/coil sheath,
use the Loop Cutter FS-5L/Q/U-1 to cut the
PolyLoop and to detach it from the device.
Refer to instructions on the reverse side for more details.
2
3
4
Note
Instructions for using PolyLoop Ligating Device:
1
Remove the plastic retainer from the PolyLoop and extend the plastic sheath until the PolyLoop is fully
covered. Insert the PolyLoop through the working channel of the endoscope.
Preparation before using PolyLoop Litigating Device:
The PolyLoop delivers a nylon loop snare designed to prevent or control bleeding following
polypectomy of pedunculated polyps.
CAUTION: This guide is only a summary of the instructions for use for the HX-400U-30 PolyLoop. Be sure to follow the Olympus EndoTherapy
PolyLoop HX-400U-30 instruction manual that was included with your HX-400U-30 purchase.
Question and Answer IconNeed help?

Do you have a question about the Olympus PolyLoop HX-400U-30 and is the answer not in the manual?

Olympus PolyLoop HX-400U-30 Specifications

General IconGeneral
BrandOlympus
ModelPolyLoop HX-400U-30
CategoryMedical Equipment
LanguageEnglish

Summary

Preparation and Instructions for PolyLoop Ligating Device

Preparation Before Use

Remove plastic retainer and extend plastic sheath until PolyLoop is covered. Insert through endoscope working channel.

Extend PolyLoop

Pull yellow cylinder proximally until it stops to extend the PolyLoop from the plastic sheath.

Position and Ligate Polyp

Position PolyLoop 1-2mm above polyp base, pull slider to ligate tissue. Monitor resistance, avoid transecting stalk.

Release PolyLoop

With yellow cylinder touching handle, push slider to release the PolyLoop.

Using and Releasing PolyLoop Ligating Device

Using an Electrosurgical Snare

Look at polyp for ligation signs (blanching/engorging). Close snare 2mm above PolyLoop on stalk.

Releasing Trapped PolyLoop

Straighten endoscope/PolyLoop. Push slider to extend hook, cut sheath/wire, then cut PolyLoop with Loop Cutter.