SUCTION PUMP KV--5
6
D Ensure the exhaust at the rear of the unit is unobstructed when the unit is oper ational.
D Store and use the KV--5 within the environmental conditions described in Section 6; failure to do so may lead to
equipment malfunction or failure.
D Only use replacement parts and accessories as specified in Section 5.3. Failure to do so may lead to equipment
malfunction or failure.
2.2 During use
D Do NOT lift the suction jar by the lid, always grip with two hands on either side of the suction jar, otherwise the cap
and jar may become separated and spillage of its contents may result.
D If aspirated material has been dr awn past the microbial filter and into the pump, the unit must be removed from use
and returned to Olympus for service.
D To prevent liquid backflow within the patient connecting tube, always leave the unit running until the patient tube is
disconnected from the suction jar.
2.3 After use
D If single use suction liners are not being used, the suction jar will contain potentially infectious body fluids. Take
suitable precautions, as dictated by local hospital policy, such as use of personal protective equipment, and empty
into a suitable waste disposal system, together with an appropriate quantity of water to dilute the jar contents.
D Single use suction liners, used filters and suction tubing will contain potentially infectious material. Dispose of
carefully as dictated by local hospital policy as clinical waste for incineration.
D Always empty the suction jar before moving the equipment. Care should be taken to obtain a firm gr ip underneath
the body of the equipment before moving. The suction jar holder must never be used as a hand hold or carrying
handle.
3 INSTRUCTIONS FOR USE
3.1 Place the KV--5 on a level flat surface.
3.2 Thoroughly inspect the suction jar for cracks or signs of damage and any suction tubing to be used for signs of wear or
damage. Do not use if damage is noted.
WARNING
Defects in the suction jar may cause the jar to implode when under vacuum.
NOTE
The suction jar may be used with or without single--use suction liners. Use without requires an optional suction
jar lid with integral float mechanism (p/no K7503760) .
3.3 Use with single- use suction liners
NOTE
The standard set supplied is a 2 litre suction system. If a 1 litre system is required, a 1 litre suction jar, a 1 litre
single --use suction liner and 340mm filter connecting tube will be r equired. These items are optional and are
detailed in Section 1.3.
(1) Refer to the instructions supplied with the single use suction liner and fit the liner to the suction jar.
NOTE
The single --use liners incorporate a solidifier to aid disposal.