EasyManua.ls Logo

Olympus KV-5 - Operating Precautions; Before Use Precautions

Olympus KV-5
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SUCTION PUMP KV--5
5
2 OPERATING PRECAUTIONS
2.1 Before use
D The unit must be used solely for its designed purpose, that is endoscopic aspiration and general medical or surgical
suction in a health care facility. This device is a ‘high vacuum, high flow’ device and is not intended for thoracic
drainage.
D Explosion hazard -- never install or use the KV--5 within the zone of risk of flammable gases.
D The unit is fitted with two M6 transit screws to protect the pump during transportation. Before use, r emove both
screws with fibre washers using the 4mm A/F Allen key provided (Figure 2-1), then remove the label from the top of
the unit. Store these items safely as they must be refitted when transporting the unit.
FIGURE 2-1
CAUTION: REMOVE TRANSIT
SCREWS BEFORE USE. REPLACE
BEFORE TRANSPORTATION.
D Excessive frothing in the suction jar during use may inhibit operation of the fluid cut--off mechanism -- use a
proprietary anti--foam agent as necessary to minimise frothing.
D The electrical installation of the room where the unit is to be used must comply with local or national regulations.
D The KV-5 is not intended to be used when positioned on inclined sur faces as the suction jar fluid cut--off mechanism
may not function correctly.
D The unit is only to be used by suitably qualified personnel in accordance with these instructions.
D Replace the microbial filter daily or when contaminated.
D Do not use anti-static tubing to connect the endoscope to the suction pump as patient safety may be compromised.
D The suction jar must never be used ‘free standing’ on the workstation or any surface. The 1 litre suction jar simply
slides into the bracket provided on the front panel of the unit as shown on page 2, the 2 litre suction jar should be
secured to workstation using the suction jar br acket.
D Periodically inspect the suction jar for obvious signs of wear/damage and replace if necessary.
D If the suction jar is dropped, it must be thoroughly inspected to ensure there are no cracks and tested to ensure the
lid is sealing correctly. If damage or a leak is evident, discard the suction jar and replace.
D When a suction jar is attached to a workstation using the optional suction jar br acket, remove the suction jar before
manoeuvring the workstation.

Related product manuals