SUCTION PUMP KV--5
5
2 OPERATING PRECAUTIONS
2.1 Before use
D The unit must be used solely for its designed purpose, that is endoscopic aspiration and general medical or surgical
suction in a health care facility. This device is a ‘high vacuum, high flow’ device and is not intended for thoracic
drainage.
D Explosion hazard -- never install or use the KV--5 within the zone of risk of flammable gases.
D The unit is fitted with two M6 transit screws to protect the pump during transportation. Before use, r emove both
screws with fibre washers using the 4mm A/F Allen key provided (Figure 2-1), then remove the label from the top of
the unit. Store these items safely as they must be refitted when transporting the unit.
FIGURE 2-1
CAUTION: REMOVE TRANSIT
SCREWS BEFORE USE. REPLACE
BEFORE TRANSPORTATION.
D Excessive frothing in the suction jar during use may inhibit operation of the fluid cut--off mechanism -- use a
proprietary anti--foam agent as necessary to minimise frothing.
D The electrical installation of the room where the unit is to be used must comply with local or national regulations.
D The KV-5 is not intended to be used when positioned on inclined sur faces as the suction jar fluid cut--off mechanism
may not function correctly.
D The unit is only to be used by suitably qualified personnel in accordance with these instructions.
D Replace the microbial filter daily or when contaminated.
D Do not use anti-static tubing to connect the endoscope to the suction pump as patient safety may be compromised.
D The suction jar must never be used ‘free standing’ on the workstation or any surface. The 1 litre suction jar simply
slides into the bracket provided on the front panel of the unit as shown on page 2, the 2 litre suction jar should be
secured to workstation using the suction jar br acket.
D Periodically inspect the suction jar for obvious signs of wear/damage and replace if necessary.
D If the suction jar is dropped, it must be thoroughly inspected to ensure there are no cracks and tested to ensure the
lid is sealing correctly. If damage or a leak is evident, discard the suction jar and replace.
D When a suction jar is attached to a workstation using the optional suction jar br acket, remove the suction jar before
manoeuvring the workstation.