FLUSHING PUMP OFP
12
7 SPARES & ACCESSORIES
NOTE
To maintain standards compliance, use only the following Spar es and Accessories with this product.
7.1 Spares
Instructions 5070135...............................
Fluid container 7501352............................
Fluid container strap 7501353........................
Irrigation tube (pack of 10) 7501669....................
Filter (pack of 10) 7501355..........................
Footswitch 7501357...............................
Fuse T1A ( pack of 5) 7501356.......................
Fuse T2A ( pack of 5) 7501404.......................
Power cable (UK) 7145454..........................
Power cable (USA) 7318766.........................
Power cable (Europe) 7145462.......................
Power cable (RoW) 7318561.........................
MAJ--855 endoscope tube 1113054 (UK part number only)....................
7.2 Accessories
Remote control cable MAJ--920
8 SPECIFICATIONS
Items Specifications
Product name Olympus Flushing Pump OFP
Standards
compliance
This symbol on the 110 --120V version indicates it is listed by
Intertek Testing Services as meeting the requirements of UL
2601--1 and CAN/CSA C22.2 601.1--M90.
3024005
C
l
a
s
s
i
f
i
c
a
t
i
o
n
Electro-magnetic
compatibility
This product complies with the requirements of EN IEC 60601-1-2 for emissions and
immunity, and as such, its operation is unlikely to be affected by, or cause interference
with, equipment meeting appropriate EMC standards. As a precaution, equipment which
may be sensitive to interference outside the limits specified by EN IEC 60601--1--2 should
not be placed in close proximity to the OFP.
C
l
a
s
s
i
f
i
c
a
t
i
o
n
(electromedical
equipment)
Type of protection
against electrical
shock
220-240V model
Classification according to EN IEC 60601-1: Class I Type BF.
110-120V model
Classification according to UL 2601-1: Class I T y pe BF
Degree of protection
against electrical
shock
In accordance with EN IEC 60601-1 and UL 2601--1, the OFP is marked with the
symbol to indicate the provision of an adequate degree of protection against electrical
shock and that it has an applied part isolated from all other parts of the equipment.
Degree of protection
against explosion
None: the Olympus OFP must NOT be used in the zone of risk of flammable anaesthetic
gases.
Mode of operation Continuous