4.4 Inspection of illumination light
33
OME-L200 INSTRUCTION MANUAL
Ch.4
4.4 Inspection of illumination light
1 Confirm that the illumination light is emitted
from the illumination lens of the surgical
microscope.
Figure 4.4
2 Referring to the instruction manual for the
surgical microscope, tap and hold the “Lamp”
button on the touch panel of the surgical
microscope for about a second and confirm
that the illumination lamp is extinguished.
Figure 4.5
3 Referring to the instruction manual for the surgical microscope, tap the “Lamp” button
on the touch panel of the surgical microscope and confirm that the illumination lamp
lights up.
4 Confirm that no error is displayed on the observation monitor or touch panel of the
surgical microscope.