App.
EMC information
59
TB-0545PC/IC/FC, TB-0535PC/IC/FC, TB-0520IC, TB-0510IC
INSTRUCTION MANUAL
Cautions and recommended electromagnetic environment
regarding portable and mobile RF communications equipment,
such as cellular phones
• Where “P” is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer, and “d” is the recommended separation
distance in meters (m).
• This instrument complies with the requirements of IEC 60601-1-2: 2001. However,
under an electromagnetic environment that exceeds its noise level,
electromagnetic interference may occur on this instrument.
• Electromagnetic interference may occur on this instrument near a high-frequency
electrosurgical equipment and/or other equipment marked with the following
symbol:
Immunity test
IEC 60601-1-2
test level
Compliance
level
Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
(150 kHz – 80 MHz)
3V (V
1
) Formula for recommended separation distance
(V
1
=3 according to the compliance level)
Radiated RF
IEC 61000-4-3
3V/m
(80 MHz – 2.5 GHz)
3 V/m (E
1
) Formula for recommended separation distance
(E
1
=3 according to the compliance level)
80 MHz – 800 MHz
800 MHz – 2.5 GHz