3.4 Preparation and inspection of compatible electrosurgical generator and ultrasonic
generator
TB-0545PC/IC/FC, TB-0535PC/IC/FC, TB-0520IC, TB-0510IC
INSTRUCTION MANUAL
Ch.3
WARNING
Always use the THUNDERBEAT in combination with the compatible electrosurgical
generator and the ultrasonic generator. The THUNDERBEAT cannot be combined
with other generators.
Table 3.1
Prepare and inspect the generators according to their instruction manuals. Should any irregularity be
observed, do not use them; contact Olympus.
Connecting/disconnecting the transducer plug to/from the
ultrasonic generator
WARNING
Do not connect the transducer plug to the ultrasonic generator while the transducer,
transducer cord and/or transducer plug are wet. Otherwise, an electric shock
and/or burns may result.
CAUTION
Always hold the transducer plug when connecting or disconnecting the transducer
plug to the ultrasonic generator. Holding a part other than the plug may
unexpectedly bend, stretch, twist, or squeeze the transducer cord, and result in
snapping of the wires.
Do not touch the transducer plug contacts. Static electricity that was accumulated
during autoclaving may cause an electric shock.
If a liquid or foreign object gets inside the transducer plug, withdraw the object by
referring to the instruction manual for the transducer. Otherwise, equipment
malfunction may result.
3.4 Preparation and inspection of compatible
electrosurgical generator and ultrasonic
generator
Ultrasonic Generator Electrosurgical Generator
USG-400 ESG-400
3.5 Connection to ultrasonic generator