EasyManua.ls Logo

Olympus THUNDERBEAT TB-0545PC - Page 64

Olympus THUNDERBEAT TB-0545PC
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
App.
60
EMC information
TB-0545PC/IC/FC, TB-0535PC/IC/FC, TB-0520IC, TB-0510IC
INSTRUCTION MANUAL
Recommended separation distance between portable and mobile
RF communications equipment and this instrument
NOTE
The guidance may not apply in some situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Portable and mobile RF communications equipment such as cellular phones
should be used no closer to any part of this instrument, including cables than the
recommended separation distance calculated from the equation applicable to the
frequency of the transmitter.
Rated maximum output
power of transmitter
P (W)
Separation distance according to frequency of transmitter (m)
(calculated as V
1
=3 and E
1
=3)
150 kHz – 80 MHz 80 MHz – 800 MHz 800 MHz – 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
d
1.2
P
=
d
1.2
P
=
d
2.3
P
=