36
appears or a
measurement
cannot be complete
after the arm cuff
inflates.
You move or talk during
a measurement and the
arm cuff does not inflate
sufficiently.
Remain still and do not
talk during a
measurement. If “E2”
appears repeatedly,
inflate the arm cuff
manually until the
systolic pressure is 30 to
40 mmHg above your
previous readings. Refer
to sub-section 3.1.
The systolic pressure is
above 210 mmHg and a
measurement cannot be
taken.
Inflate the arm cuff
manually until the
systolic pressure is 30 to
40 mmHg above your
previous readings. Refer
to sub-section 3.1.
appears
The arm cuff is inflated
exceeding the maximum
allowable pressure.
Do not touch the arm cuff
and/or bend the air tube
while taking a
measurement.
If inflating the arm cuff
manually, refer to the
end of sub-section 3.1.
appears
You move or talk during
a measurement.
Vibrations disrupt a
measurement.
Remain still and do not
talk during a
measurement.
Display/Problem Possible Cause Solution

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron BP7250 and is the answer not in the manual?

Omron BP7250 Specifications

General IconGeneral
Model NumberBP7250
TypeUpper Arm Blood Pressure Monitor
Bluetooth ConnectivityYes
Irregular Heartbeat DetectionYes
User Profiles2
Measurement MethodOscillometric
Power Source4 AA batteries or AC adapter
Average ReadingAverages 3 readings taken within 10 minutes
Cuff Size9 to 17 inches
WeightApprox. 400g (without batteries)