32
9. Guidance and Manufacturer’s Declaration
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BP785N including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of 
this OMRON BP785N including AC-adapter should assure that it is used in such environment.
Immunity test IEC 60601 test level
Compliance 
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V rms
3 V/m
Portable and mobile RF communications equipment should be used no closer to 
any part of the OMRON BP785N including AC-adapter including cables, than the 
recommended separation distance calculated from the equation appropriate to the 
frequency of the transmitter.
Recommend separation distance
d = 1.2 √P
d = 1.2 √P 
80 MHz to 800 MHz
d = 2.3 √P 
800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) 
according to the transmitter manufacturer and d is the recommended separation 
distance in meters (m).
Field strengths from fixed RF transmitters as determined by an electromagnetic 
site survey,
a
 should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following 
symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, 
objects, and people.
a
 Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radio, AM and FM radio 
broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the OMRON 
BP785N including AC-adapter is used exceeds the applicable RF compliance level above, the OMRON BP785N including AC-adapter should 
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or 
relocating the OMRON BP785N including AC-adapter.
b
 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
HEM-7321-Z_C_M.book  Page 32  Monday, June 16, 2014  12:09 PM