EasyManua.ls Logo

Omron CV500 - Page 96

Omron CV500
183 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
83
Fault Check Flowchart
A
C
B
B
A
B
RUN
indicator not lit
NO
RUN or
MONITOR
mode?
YES
ERROR indi-
cator lights?
NO
Can Online
mode be set?
YES
NO
WDT indica
-
tor lights?
NO
YES
Can Online
mode be set?
YES
NO
WDT indica
-
tor lights?
YES
ON
A30600 ON?
OFF
ON
A30601 ON?
OFF
ON
A30602 ON?
OFF
ON
A30603 ON?
OFF
NO
RUN indica-
tor lights?
YES
*When an I/O verifica
-
tion error is specified
as a fatal error in the
PC Setup.
**This error does not
occur when disabled
as an error in the PC
Setup.
At this time, it is assumed that the POWER indicator is normally
lit.
(T
o next page)
YES NO
STAR
T input OFF
Connect peripheral de
-
vice and check error
Check word A306
I/O verify error*
SYSMAC BUS terminator
missing
SYSMAC BUS/2 termina
-
tor missing**
Close STAR
T input termi
-
nals of Power Supply Unit
Unit is disconnected. Iden
-
tify erroneous Unit by veri
-
fying I/O table, mount Unit
correctly
, and then create
I/O table
Check terminator and
Slave Rack Power Supply
Unit of SYSMAC BUS
Check terminator and
Slave Rack Power Supply
Unit of SYSMAC BUS/2
End
Replace Unit
T
urn power OFF and ON
again
Set RUN or MONITOR
mode
Troubleshooting Flowcharts Section 5-2

Other manuals for Omron CV500

Related product manuals