31
8. SPECIFICATIONS
Model
BP7000 HEM-7600T-Z
Display OLED display
Cuff pressure range Pressure: 0 to 299 mmHg
Measurement range Pressure: 40 to 260 mmHg
Pulse: 40 to 180 beats / min.
Accuracy Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic rapid deflation
Measurement method Oscillometric method
Transmission method
Bluetooth
low energy technology
Wireless communication Frequency range : 2.4 GHz (2400 - 2483.5 MHz)
Modulation : GFSK
Effective radiated power : <20 dBm
IP classification IP 22
Power source
4 “AAA” batteries 1.5V
Battery life Approximately 300 measurements
(using new alkaline batteries)
Durable period (Service life) Monitor : 5 years
Operating conditions 50°F to 104°F (10°C to 40°C) /
15 to 90% RH (non-condensing) / 800 to 1060 hPa
Storage / Transport
conditions
-4°F to 140°F (-20°C to 60°C) /
10 to 90% RH (non-condensing)
Weight Monitor : Approximately 8 1/2 oz. (240 g)
not including batteries
Dimensions Monitor : Approximately 3 1/2" (w) × 4 5/8" (h) × 1" (l)
(85 mm × 120 mm × 20 mm)
(not including the arm cuff)
Arm circumference 9" to 17" (22 to 42 cm)
Contents Monitor, 4 “AAA” batteries, instruction manual, quick start
guide
Applied part Type BF
Protection against electric
shock
Internally powered ME equipment
®

Related product manuals