32
Power supply
Environmental Conditions
Non-Invasive Blood Pressure (NIBP)
* Comparison with auscultation method performed by a trained professional.
DIA determined by the auscultation method is "K5".
AC adapter Input voltage range: AC 100 V to 240 V
Rated Current: 1.6 A
Frequency: 50/60 Hz
Output voltage range: DC 6 V ±5%
Optional rechargeable battery Type: 3.6 V, 1900 mAh
Number of operation cycles when fully charged: 300
Measurement conditions
New battery fully charged
Ambient temperature of 23°C (73.4°F)
Using M-size cuff
SYS 120 / DIA 80 / PR 60 (Inflation setting: AUTO)
One 5-minute cycle consisting of "cuff measurement time + wait time"
Operational temperature and
humidity
Temperature range: 5 to 40°C (41 to 104°F)
Humidity range:
15 to 85%RH (not condensed)
Atmospheric pressure: 700 to 1060hPa
Storage and transportation Temperature range: -20 to 60°C (-4 to 140°F)
Humidity range: 10 to 95%RH (not condensed)
Atmospheric pressure: 500 to 1060hPa
EMC: Reference standard IEC60601-1-2:2007
Medical electrical equipment -
Part 1-2: General requirements for basic safety and essential performance
- Collateral standard: Electromagnetic compatibility - Requirements and
tests.
Measurement technology Oscillometric
Measurement method Dynamic Linear Deflation method
Pressure display range 0 to 300 mmHg
Pressure display accuracy Within ±3 mmHg or 2 %
NIBP measurement range SYS 60 to 250 mmHg
DIA 40 to 200 mmHg
Pulse rate 40 to 200 /min
NIBP accuracy* Maximum mean error within ±5mmHg
Maximum standard deviation within ±8mmHg
Pulse rate accuracy Within ±5 % of reading
Reference standard: IEC 80601-2-30:2009
ANSI/AAMI SP-10:2002+A1:2003+A2: 2006/(R)2008
ISO81060-2:2009
HBP-1300.book Page 32 Wednesday, March 7, 2012 9:28 AM