EasyManuals Logo

Omron HEM-432C User Manual

Omron HEM-432C
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
USING THE MEMORY FUNCTION
16
Every time you complete a measurement the monitor stores the blood
pressure and pulse rate in the memory. The Blood Pressure Monitor
automatically stores up to 30 measurement values. When 30 sets of
measurement values are stored in the memory, the oldest record is
deleted to save the most recent measurement values.
AVERAGING FUNCTION
The monitor also calculates an average reading based on the values of
the three most recent measurements that were taken within 10 minutes
of the most recent reading.
NOTES:
If only two readings are stored in the memory for the 10-minute
period, the average is based on the two readings.
If one reading is stored in the memory for the 10-minute period,
this reading is displayed as the average.
If the date and time is not set, the measurement values and the average
value will not be stored in the memory correctly.
TO DISPLAY THE AVERAGE VALUE
Average Value Symbol
1.
Press the ON/OFF button to turn the power on.
2.
Press the MEMORY button.
The average value symbol will display on the screen with the
average reading.
FINAL.HEM-432CN2-IM 2/23/10 10:56 AM Page 16

Other manuals for Omron HEM-432C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron HEM-432C and is the answer not in the manual?

Omron HEM-432C Specifications

Measurement Specifications IconMeasurement Specifications
pressure measurement range0 - 280 mmHg
pulse measurement range40 - 200/minute
pressure accuracy±3 mmHg or 2 % of reading
pulse accuracy±5% of reading
Operating Environment IconOperating Environment
operating temperature50°F to 104°F (10°C to 40°C)
operating humidity30 to 85% RH maximum
storage temperature-4°F to 140°F (-20°C to 60°C)
storage humidity10 to 95% RH maximum
Power IconPower
power source4 “AAA” batteries
battery usageApprox. 300 usages
Weight and Dimensions IconWeight and Dimensions
unit weight1.0 lbs
unit dimensions5.13" x 3.5" x 6.88"
master weight14.25 lbs
master dimensions22" x 11" x 8.25"
monitor weightApproximately 8 oz (225 g)
cuff sizeApproximately 5 1/2" x 18 7/8" (140 mm x 480 mm)
cuff circumference9" to 13" (229 mm x 330 mm)