23
Air plug is disconnected.
Insert the plug securely.
Refer to section 3.1.
Arm cuff is applied too
loosely.
Apply the arm cuff tighter.
Refer to section 3.1.
Air is leaking from the arm
cuff.
Replace the arm cuff with a new
one.
Refer to section 5.3.
The cuff may not be the
correct size for your arm.
Measure your arm circumference
to make sure that you are using
the correct size cuff.
Refer to section 5.3.
Movement during
measurement and the arm
cuff has not been inflated
sufficiently.
Repeat measurement. Remain
still and do not talk during
measurement.
Refer to section 3.3.
If “E2” appears repeatedly,
inflate the arm cuff manually until
it is 30 to 40 mmHg above your
previous measurement result.
Refer to section 3.3.
The arm cuff was inflated
exceeding the maximum
allowable pressure, and
then deflated automatically.
Do not touch the arm cuff and/or
bend the air tube while taking a
measurement. Do not inflate the
arm cuff more than necessary.
Refer to section 3.3.
Error Display Cause Solution

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron HEM-9210T and is the answer not in the manual?

Omron HEM-9210T Specifications

General IconGeneral
BrandOmron
ModelHEM-9210T
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals