6. Specifications
Product description Automatic Upper Arm Blood Pressure Monitor
Product category Electronic Sphygmomanometers
Model (code) M1 Basic (HEM-7121J-AF) Display LCD digital display
Cuff pressure range 0 to 299mmHg
Pulse measurement range
40 to 180 beats / min.
Blood pressure measurement range
20 to 280mmHg
Accuracy Pressure: ±3mmHg / Pulse: ±5% of display reading
Inflation Automatic by electric pump Deflation
Automatic pressure release valve
Measurement method Oscillometric method Operating mode Continuous operation
IP classification Monitor: IP20 / Optional AC adapter: IP21 (HHP-CM01) or IP22 (HHP-BFH01)
Rating DC6 V 4.0 W Applied part Type BF (arm cuff)
Power source 4 “AA” batteries 1.5V or optional AC adapter
(INPUT AC 100 - 240V 50-60Hz 0.12 - 0.065A)
Battery life Approximately 1000 measurements (using new alkaline batteries)
Durable period (Service life) Monitor: 5 years / Cuff: 5 years / Optional AC adapter: 5 years
Operating conditions +10 to +40°C / 15 to 90% RH (non-condensing) / 800 to 1060hPa
Storage / Transport conditions -20 to +60°C / 10 to 90% RH (non-condensing)
Contents
Monitor, arm cuff (HEM-CR24), 4 “AA” batteries, Instruction Manual and
Protection against electric shock Internally powered ME equipment (when using only batteries)
Class II ME equipment (optional AC adapter)
Weight Monitor: approximately 250g (not including batteries) / Arm cuff: approximately 130g
Dimensions (approximately value) Monitor: 108mm (W) × 83mm (H) × 140mm (L)
Arm cuff: 145mm × 466mm (air tube: 610mm)
Cuff circumference applicable to
the monitor
170 to 420 mm
(included arm cuff: (220 to 320 mm))
Maximum temperature
of the applied part
Lower than 48 °C
Memory Last measurement
Note
• These specifications are subject to change without notice.
• This monitor is clinically investigated according to the requirements of EN ISO 81060-2:2014 and complies with EN ISO 81060-2:2014 and
ENISO 81060-2:2019 + A1:2020 (excluding pregnant and pre-eclampsia patients). In the clinical validation study, K5 was used on 85 subjects for
determination of diastolic blood pressure.
• IP classification is degrees of protection provided by enclosures in accordance with IEC 60529. This monitor and optional AC adapter are
protected against solid foreign objects of 12.5 mm diameter and greater such as a finger. The optional AC adapter HHP-CM01 is protected
against vertically falling water drops which may cause issues during a normal operation. The optional AC adapter HHP-BFH01 is protected
against oblique falling water drops which may cause issues during a normal operation.
• Operating mode is classification in accordance with IEC 60601-1.