3. Error Messages and Troubleshooting
If any of the below problems occur during measurement, check to make sure that no other electrical device is within 30 cm. If the problem persists,
please refer to the table below.
Display/Problem Possible Cause Solution
appears or the
arm cuff does not
inflate.
The [START/STOP] button was pressed
while the arm cuff is not applied.
Press the [START/STOP] button again to turn the monitor
off. After inserting the air plug securely and applying the
arm cuff correctly, press the [START/STOP] button.
Air plug is not completely plugged
into the monitor.
Insert the air plug securely.
The arm cuff is not applied correctly.
Apply the arm cuff correctly, then take another
measurement. Refer to section 5 of instruction manual
.
Air is leaking from the arm cuff. Replace the arm cuff to the new one. Refer to section 12
of instruction manual
.
appears or a
measurement
cannot be
completed after the
arm cuff inflates.
You move or talk during a
measurement and the arm cuff does
not inflate sufficiently.
Remain still and do not talk during a measurement. If
“E2” appears repeatedly, inflate the arm cuff manually
until the systolic pressure is 30 to 40 mmHg above your
previous readings. Refer to section 11 of instruction
manual
.
Due to the systolic pressure is above
210mmHg, a measurement cannot
be taken.
appears
The arm cuff is inflated exceeding the
maximum allowable pressure.
Do not touch the arm cuff and/or bend the air tube while
taking a measurement. If inflating the arm cuff manually,
refer to section 11 of instruction manual
.
appears
You move or talk during a
measurement. Vibrations disrupt a
measurement.
Remain still and do not talk during a measurement.
appears
The pulse rate is not detected
correctly.
Apply the arm cuff correctly, then take another
measurement. Refer to section 5 of instruction manual
.
Remain still and sit correctly during a measurement.
If the “ symbol continues to appear, we recommend
you to consult with your physician.
/ /
appears
does not flash during
a measurement

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron M3 HEM-7154-E and is the answer not in the manual?

Omron M3 HEM-7154-E Specifications

General IconGeneral
TypeAutomatic
Cuff size220 - 420 cm
Control typeButtons
Display typeLCD
Product colorBlue, White
Number of users2 user(s)
Memory registers120
Pulse rate range40 - 180
Placement supportedUpper arm
Units of measurementmmHg
Mean arterial pressure-
Systolic blood pressureYes
Pressure accuracy6 mmHg
Pressure measurement range0 - 299 mmHg
Battery typeAA
Battery voltage1.5 V
Battery life (CIPA standard)1000 shots
Number of batteries supported4
Weight and Dimensions IconWeight and Dimensions
Depth153 mm
Width105 mm
Height87 mm
Weight310 g