4. Troubleshooting and Maintenance
12
Forehead measurement mode: The temperature
taken is higher than +42.2°C (108°F).
Surface measurement mode: The temperature
taken is higher than +80°C (176°F).
Use the unit within its measurable range. If a
malfunction still exists, please contact an OMRON
representative.
Forehead measurement mode: The temperature
taken is lower than +34°C (93.2°F).
Surface measurement mode: The temperature
taken is lower than -22°C (-7.6°F).
Use the unit within its measurable range. If a
malfunction still exists, please contact an OMRON
representative.
The unit cannot be powered on to ready stage.
Change to a new battery.
(Refer to section 4.4)
Error Display Cause Remedy
MC-720-AP_A_M.book Page 12 Friday, May 22, 2015 11:01 AM