7
SETTING UP YOUR METER
CAUTION: Matching the code on the meter and the code on the test strip vial is essential to obtaining accurate
results. Each time you test, check to make sure the code numbers match.
Insert a test strip to turn on the meter
Remove a test strip from its vial. With clean, dry hands, you may touch the test strip anywhere on
its surface. Do Not bend, cut or modify the test strips in any way. Use each test strip immediately
after removing it from the vial. Insert the test strip into the test port as shown, with the three
contact bars facing you. Push the test strip in as far as it will go.
After the black start-up screen appears, the meter will display the code from your last
test. If a ashing “– –” appears instead of a code number, such as when you are rst
using the meter, follow the instructions in step 3 to change to a numerical code.
Match the code on the meter with the code on the test strip vial
If the code on the meter does not match the code on the test strip vial, press
or
to match
the code number on the test strip vial. The new code number will ash on the display for three
seconds, then briey stop ashing, after which the display will advance to the APPLY BLOOD
screen.
If the codes already match, press
to go to the APPLY BLOOD screen. When you do not
make a change after ve seconds, the display will advance to the APPLY BLOOD screen.
The meter is now ready to perform a blood glucose test.
NOTE:
• If the APPLY BLOOD screen appears before you are sure the codes match, remove the test strip, and re-start from step 1.
• If you change APPLY BLOOD to APPLY CONTROL by mistake, press
to change it back to APPLY BLOOD.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the OneTouch ULTRA2 and is the answer not in the manual?

OneTouch ULTRA2 Specifications

Power IconPower
meter power sourceOne replaceable 3.0 Volt CR 2032 lithium battery
backlight power sourceOne replaceable 3.0 Volt CR 2032 lithium battery
battery ratings2 x 3.0 V d.c., 60 mA
Operating Environment IconOperating Environment
temperature range6–44°C
relative humidity range10–90%
altitude rangeup to 3048 metres
Weight and Dimensions IconWeight and Dimensions
weight42.5 grams
size7.92 x 5.72 x 2.29 cm