EasyManua.ls Logo

OneTouch Verio Flex - Regression Statistics

OneTouch Verio Flex
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
103
Detailed information about your system
9
Regression Statistics
Samples were tested in duplicate on each of three test
strip lots. Results indicate that the OneTouchVerioFlex™
System compares well with a laboratory method.
# of Subjects # of Tests Slope Intercept
(mmol/L)
100 600 1.01 -0.02
95% CI
Slope
95% CI
Intercept
(mmol/L)
Std. Error
(S
y.x
)
(mmol/L)
R
2
1.00 to 1.01 -0.12 to 0.08 0.64 0.99