User Performance Accuracy
A study evaluating glucose values from fingertip capillary
blood samples obtained by 167 lay persons showed the
following results:
93.1% within ±0.83mmol/L of the medical laboratory
values at glucose concentrations below 5.55mmol/L, and
98.6% within ±15% of the medical laboratory values at
glucose concentrations at or above 5.55mmol/L.
97.6% of the total number of samples were within
±0.83mmol/L or ±15% of the medical laboratory values.