Detailed information about your system
9
Precision
Within Run Precision (300 Venous Blood Samples Tested
per Glucose Level)
Data generated using the OneTouchVerioFlex™ Meter.
Target
Glucose
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coecient
of Variation
(%)
2.2 2.0 0.05 2.26
5.0 4.9 0.10 1.95
7.2 7.1 0.14 1.95
11.1 11.1 0.22 1.98
19.4 19.1 0.34 1.79
Results show that the greatest variability observed
between test strips when tested with blood is 0.10mmol/L
SD or less at glucose levels less than 5.55mmol/L, or
1.98% CV or less at glucose levels at 5.55mmol/L or above.