 8 Detailed information
about your system
Comparing meter results to laboratory
results
Results obtained from the OneTouch®Verio® Meter
and laboratory tests are reported in plasma-equivalent
units. However, your meter result may differ from your
lab result due to normal variation. A result from your
OneTouch®Verio® Meter is considered accurate when it is
within ±20% of the lab result.
Meter results can be affected by factors that do not affect
lab results in the same way, which may cause a difference
of more than ±20%. Specific factors that may cause your
meter result to vary from your lab result by more than
±20% include:
• You have eaten recently. This can cause a result from
fingertip testing to be up to 3.9mmol/L higher than a
lab test using blood drawn from a vein.
1
• Your haematocrit is above 60% or below 20%.
• You are severely dehydrated.
• For additional information, refer to the
OneTouch®Verio® Test Strip Insert.
1
Sacks, D.B.: "Carbohydrates." Burtis, C.A., and Ashwood
E.R. (ed.), Tietz Textbook of Clinical Chemistry,
Philadelphia: W.B. Saunders Company (1994), 959.