System Accuracy
Diabetes experts have suggested that glucose meters
should agree within 0.83mmol/L of a laboratory method
when the glucose concentration is <4.2mmol/L, and
within 20% of a laboratory method when the glucose
concentration is ≥4.2mmol/L or higher. Samples from
100 patients were tested using both the OneTouch®Verio®
System and the YSI 2300 Glucose Analyzer laboratory
instrument.
System Accuracy Results for Glucose
Concentrations <4.2 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within
±0.28 mmol/L
Within
±0.56 mmol/L
Within
±0.83 mmol/L
50.7%
(70/138)
92.0%
(127/138)
100%
(138/138)