Detailed information
about your system
8
118
System Accuracy Results for Glucose
Concentrations ≥4.2 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within ±5% Within ±10% Within ±15% Within ±20%
63.0%
(291/462)
90.0%
(416/462)
97.4%
(450/462)
99.6%
(460/462)
System Accuracy Results across the entire
Glucose Range
Percent (and number) of meter results that match the
laboratory test
Within ±0.83 mmol/L or ±20%
99.7% (598/600)
Therefore, 99.7% of the total results obtained with the
OneTouch®Verio® System achieved the goal suggested by
the diabetes experts.
However, experts have suggested greater accuracy of
Blood Glucose Monitoring Systems is desirable. The
OneTouch®Verio® Blood Glucose Monitoring System
has demonstrated accuracy within 0.83mmol/L of the
laboratory method for glucose concentrations lower than
5.55mmol/L, and within 15% of the laboratory method for
glucose concentrations 5.55mmol/L or higher.