GFPIII operation manual Opti-Sciences, Inc.
21
Data formats:
The format is stored in a comma separated variable format that can be easily imported into popular
spreadsheet programs. A header in each file identifies the specific data field.
Sample Test Format:
Sample #, Date, Time, Raw Signal, CFR, Chlorophyll content in mg m
-2
1,Oct/11/2011,11:49:43,751,0.33, 600
2,Oct/11/2011,11:50:02,796,0.35, 613
Specifications:
Source:
Solid state modulated LED device providing excitation wavelengths from 460 nm with a half
bandwidth of 15 nm
Detector:
Two solid state detectors . One filtered for detection from 698nm to 708 nm and the other
filtered for a range from 730nm to 740nm.
Noise:
Typically < 2% of full scale, but is dependant upon calibrated range
Sensitivity:
signal integration over 5 seconds.
Sensing probe:
Fiber has 0.10” diameter active region contained within a 0.15" diameter stainless steel tip
Calibration:
Solid state sensors and light sources provide no significant variation over the life of the
instrument. The measurement represents a ratio of fluorescence at two different emission
wavelengths. The research shows that this ratio provides a precise measurement of Chlorophyll
content from 41mg m-2 to 675 mg m-2 using Beech, Elm or Wild Vine. Calibration may be
checked by comparing chemical tests to read-out values. Parameters in the direct chlorophyll
content formula may be changed to conform with chemical test results for any unexpected non-
conforming species. The values and equation in the instrument conform with the equation
supplied by Gitleson (1999).
Repeatability is dependent on the emission signal strength. Measurements in the yellow signal
strength range may require multiple measurements and averaging for more reliable results.
User Interface:
320x480 color TFT LCD screen with touch panel surface for user input
Data capacity:
Internal, non volatile 2GB memory for storage of calibration and data files