CCM-200 plus
Chlorophyll Content Meter
Question and Answer IconNeed help?

Do you have a question about the opti-sciences ccm-200 plus and is the answer not in the manual?

opti-sciences ccm-200 plus Specifications

General IconGeneral
Brandopti-sciences
Modelccm-200 plus
CategoryMeasuring Instruments
LanguageEnglish

Summary

Applications & Features

How it Works

Measurement Principle

Explains how the CCM-200 plus uses transmittance at 931nm and 653nm to calculate the Chlorophyll Content Index (CCI).

System Settings

Measurement & Operation

Measurements

Calibration

Zeroing the Instrument

Explains that calibration (zeroing) is required every time the unit is turned on to check the chamber path.

Test Types

Multi-Point averaging

Explains multi-point averaging for nitrogen management schemes, allowing 2 to 30 points to be averaged.

Multi-Point averaging with std. deviation flyer removal

Details averaging 10 to 30 samples with automatic rejection of readings outside 2 sigma.

Data Management

Technical Specifications

ERROR CODES

Boot up memory errors

Lists errors like 'SD init error', 'SD error #xx', and 'NV load error' with a fix involving battery removal.

File System Corrupt Error

Addresses 'File System Corrupt!' errors, often due to power loss or improper disconnection, and suggests formatting.

Calibration Errors

Clean Sample Chamber Error

Occurs during calibration if light through the chamber is too low, suggesting dirt or mechanical issues.

Source/Detector Fail Error

Indicates a fault with sensing components, requiring factory inspection.

Chamber Light Leak Error

Signals worn gaskets or a defective detector, requiring service for replacement.

Light Leak Warning

Indicates too much light during zero phase, possibly due to worn gaskets or non-flat samples.

Data view & output errors

File Corrupt Error

Occurs due to corrupted files from aborted transfers or improper shutdown, suggesting battery removal instead of power switch.

Technical Support

Contact Information

Provides contact details for Opti-Sciences, Inc. in North America, including address, phone, fax, internet, and email.